ChemiDoc MP with Image Lab Software
101
9 Maintenance
This chapter includes instructions for maintaining the universal hood in proper
working condition by replacing parts.
UV Transilluminator Lamp and Starter
Replacement
Note: The UV filter surface should always be kept clean from the
chemical agents used as gel dyes. Use protective gloves when
touching the UV transilluminator cover.
Depending on usage, the UV bulbs and starters last for many years. Replace
bulbs when you notice them flickering. If a bulb does not turn on when it is
new or moved, replace the bulb starter and test the bulb again.
Three types of bulbs are available. The catalog numbers are listed in Ordering
Information on page 118. The standard bulb is 302 nm. Optionally, the 254 nm
bulb is used for cross-linking of protein, and the 365 nm bulb is used to
minimize denaturing of DNA.
To replace the lamps:
1. Turn off the power.
2. Disconnect the power cord from the universal hood.
3. Remove the four screws located on the left/right sides of the
transilluminator cover.