ChemiDoc MP with Image Lab Software | Acquiring Images
26
2. Choose a unique application name.
3. Select a light source, filter, and image color on the drop-down menus.
4. Select a binning setting. Choosing a higher binning setting combines
pixels to increase the amount of signal without increasing noise. While a
higher setting provides optimal sensitivity for low-light applications
such as chemiluminescence, it also reduces image resolution.
5. Click OK to create your custom application.
6. Choose the Imaging Area—select from the list of Bio-Rad gels or
enter
image area dimensions. The red line represents the imaging area
for the selected gel, and the gray rectangle represents the imager sample
stage
.
7. Choose Image Exposure Time—if you select the top radio button,
Image Lab software acquires an image estimating an optimal exposure
time. This option ensures the best use of the dynamic range.
If you choose Intense Bands from the drop-down menu, exposure
is optimized for all bands
If you choose Faint Bands on the drop-down list, a longer exposure
will make faint bands more visible, but more prominent bands may
be overexposed
Select the bottom radio button to manually override automated
imaging that may cause saturation (overexposure) of more