ChemiDoc MP with Image Lab Software | Appendix B. Troubleshooting
110
Printout does not look
like the monitor image
Monitor settings are wrong
Printer settings are wrong
Refer to monitor instructions to correct
settings
Refer to the printer manual to correct
settings
Light leakage into the
darkroom
The lens body is not seated
properly against the gasket
on the hood’s adapter plate
Loosen the thumbscrew and seat the
lens properly against the gasket on the
hood’s adapter plate
Unable to focus on the
sample using white
light transilluminator or
conversion screen
Focus is not calibrated for
samples using this light
source
Select Edit > Instrument Setup to
recalibrate focus using the proper
accessory
Lens limits seem
artificially restricted
Camera lens not seated
properly on lens mounting
plate
Reseat camera on lens mounting plate
Problem Possible Cause Solution