ChemiDoc MP with Image Lab Software | Appendix B. Troubleshooting
110
Printout does not look
like the monitor image
• Monitor settings are wrong
• Printer settings are wrong
• Refer to monitor instructions to correct
settings
• Refer to the printer manual to correct
settings
Light leakage into the
darkroom
• The lens body is not seated
properly against the gasket
on the hood’s adapter plate
• Loosen the thumbscrew and seat the
lens properly against the gasket on the
hood’s adapter plate
Unable to focus on the
sample using white
light transilluminator or
conversion screen
• Focus is not calibrated for
samples using this light
source
• Select Edit > Instrument Setup to
recalibrate focus using the proper
accessory
Lens limits seem
artificially restricted
• Camera lens not seated
properly on lens mounting
plate
• Reseat camera on lens mounting plate