ChemiDoc MP with Image Lab Software | Analyzing Images
80
3. Select a unit of measure from the choices listed in the Units drop-down
list.
4. Select a regression method from the drop-down list.
Keep the following guidelines in mind while making your choice.
Linear—generates a straight line that is the best fit of the values you
provided and is preferred in most cases.
Point-to-point—generates a curve in which each data point is
connected directly to the next, regardless of the shape of the resulting
curve.
Cubic spline—generates a smooth curve that connects each data
point. At least four standard points are required to use this method of
least-squares polynomial fits.
5. Click Standard Curve in the toolbar, select the Absolute Quantity
Standard Curve tab, and a calibration curve displays. Standards are
Regression Method
Minimum Number of
Standard Bands
Minimum Number with
Force Through Origin
Option
Linear 2 1
Point-to-point 2 1
Cubic spline 5 4