ChemiDoc MP with Image Lab Software | Maintenance
102
4. Remove the cover with the UV glass by sliding it forward, then lifting up.
5. Place it on a nonabrasive surface so that the glass does not get
scratched or damaged.
Note: Do not put the UV cover directly on the bench. Wear gloves
when touching the lamps.
6. Rotate the lamp until it becomes loose and the pins come to a vertical
position.
7. Remove the lamp. Install the new lamp by rotating so that the pins are
horizontal and the lamp is tight.
8. Refer to the pictures above to replace a starter. Remove the starter by
rotating it counterclockwise, and then pull it out.
9. To replace a starter, insert it into the holder and rotate clockwise.
10. .Reassemble the cover and retighten the screws on both sides.