ChemiDoc MP with Image Lab Software | Analyzing Images
78
To select a reference band from an image and quantitate all other bands
relative to it:
1. Click the Select button.
2. Click the band you want to use as a reference. A small R appears near
the band you selected.
To review the relative band quantities:
Go to the Rel. Quant. column of the Analysis table (Lane and Band
tab). The relative quantity is the ratio of the band volume divided by
the reference band volume
All other bands now display numerical values that are relative to the reference
band. Values above 1.00 indicate that the band quantity is greater than the
reference band. Values below 1.00 indicate the band quantity is less than that
of the reference band.