Alpha Innotech User’s Manual, version 4
5-4
A Skewed Template
Specifying the Areas to be Measured
For microtiter plate analysis, the toolbox controls labeled Center and Outer specify the
size of the inner circle and outer torus scanned to measure each well. The number in the
center of each of these controls indicates the diameter of the circles. Click on an arrow to
increase or decrease the diameters. As the number is adjusted, the objects on the
template grow larger or smaller, reflecting the changes.
When specifying the size of the area to be measured, exclude shadows or other imaging
artifacts. Depending upon the lighting conditions when the image was captured, there
may be a crescent-shaped shadow obscuring part of the bottom of the wells, especially
those near the edge of the microtiter plate. We recommend decreasing the size of the
sampling area so that these regions are not included.
To measure the total density within each object, reduce the Center value to zero (0) and
adjust the Outer value until the objects on the screen are slightly smaller than the areas of
interest on the image. Reposition the corner objects using the mouse until the objects are
properly aligned over the objects.
Measuring Density
Once the objects are sized and positioned correctly, click on the SCAN button in the tool
box. The density values will be calculated and displayed under each object.